Helmke Drum Data Interpretation

Helmke Drum Data Interpretation 13.11.2026

What is Helmke Drum Testing?

  • A test to measure how many particles are released from a garment, or item, when it is moved.
  • It follows the method described in the US standard IEST-RP-CC003.5.
  • The garment is placed inside a rotating drum and tumbled under dry, controlled conditions.
  • A particle counter then measures the particulate in the drum air during the 10-minute tumbling and calculates an average per minute, which is used to determine the particle emission rate from the garment.
  • The particle emission rate is representative of how many particles a garment releases when worn.
  • Helmke drum measurements are not suitable for mops/wipes/gloves (these use a different test).
Handling Cleanroom Garments to place into a Helmke Drum Test An open Helmke Drum ready for receiving a cleanroom garment for testing. A closed Helmke Drum rotating whilst performing a test on a cleanroom garment.
Cleanroom garment is folded. Cleanroom garment is placed in the test drum. Particle sampling is conducted while the drum tumbles the garment.

 

How are cleanliness classifications defined?

  • The IEST standard for specific garment types defines limits for: frocks, coveralls and hoods.
  • Limits are based on medium-sized garments.
  • Based on the particulate levels, items will be assigned a ‘cleanliness category’.

What do the cleanliness categories mean in practical terms?

Cleanliness categories indicate the expected particle release under normal use:

  • Category I: Very low particle shedding = suitable for critical cleanroom areas.
  • Category II: Moderate particle shedding = suitable for less critical cleanrooms.
  • Category III: Higher particle shedding = may be used in non-critical or peripheral areas.

Why does garment or item size matter?

  • Larger items have more fabric which will release more particles.
  • Smaller items naturally release fewer particles overall.
  • Size must be considered to avoid misleading results.

How are results reported?

Method 1: Direct comparison to standard limits

  • Reports the particles from a total item and compares directly to standard limits.
  • Item size is not factored in.
  • Small items can appear to perform better as they have less area to produce particles.
  • This can give misleading results.

Method 2: Size-adjusted comparison (Micronclean’s preferred method)

  • Results and standard limits are adjusted per square metre (m2) of material/item.
  • Limits are adjusted based on the actual surface area of the item.
  • All items are therefore directly compared to the same particle emission rate standard.
  • Allows fair comparison between different sizes and item types and prevents small items from seeming to be suitable for a cleanroom just because they are small.

What should customers consider when comparing supplier data?

  • Not all suppliers use the same reporting method.
  • Always check how the category was determined.
  • Results from different methods will not be comparable.

Key takeaways

  • Helmke drum testing measures particle release under dry movement.
  • Results must be interpreted with item size in mind.
  • Size‑adjusted reporting provides the most meaningful comparison.
  • Understanding the method used is essential when comparing data.